SOMATOM Force, Definition AS, Definition Edge, Definition Flash, Drive, Confidence and Edge Plus
FDA Recall Z-0771-2021 · Posted None
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeJAK
Date Initiated2020-11-24
Date Terminated2024-02-06
Product Quantity3044
Reason for Recall
System does not trigger a cancel command followed by a reload of the scan with the new parameter(s). This issue may result in a delay in diagnosis and/or need for patient rescan
Firm Action
Customer notification letters will be sent starting December 14, 2020 to all affected customers via Customer Advisory Notice CT057/20/S. The technical solution, software versionVB20A_SP3, will be distributed to all affected customers free of charge via CT056/20/S. For questions or technical support, contact your local application specialists or your local service/sales organization at 1-800-888-7436.
Root Cause
Software design
Distribution Pattern
US Nationwide