There is a complaint regarding internal check cable guide mechanism failure for the Nucletron Flexitron brachytherapy afterloader system. Due to an obstruction during the check cable out drive, the internal guiding tube became displaced and the check cable became damaged. This issue could cause treatment interruption or in a very unlikely scenario, incorrect source positioning and radiation treatment errors.
FDA Recall Z-0772-2020 · Posted None
Advertisement
Recalling FirmNucletron BV
Recall StatusTerminated
Product CodeJAQ
Date Initiated2019-11-18
Date Terminated2023-03-28
Product Quantity383
Reason for Recall
A component failure may lead to treatment interruption or incorrect source positioning.
Firm Action
The recalling firm will send a Field Safety Notice 806-01-BTD-001 to all affected customers. The notice informs user of the specific product and version numbers affected by the issue, and any work arounds that can be used to avoid the issue. A copy of the Field Safety Notice is to be kept with the most current labeling and all personnel working with the product should be made aware of the content of the letter. Customers are instructed to complete and return the Field Safety Notice Acknowledgement form to their local Eleka representative as soon as possible and no longer than 30 days from receipt.
Root Cause
Process control
Distribution Pattern
Distribution in United States and worldwide