Pilling Wecksorb Cylindrical Sponges
FDA Recall Z-0772-2023 · Posted 2022-12-28
Advertisement
Recalling FirmTELEFLEX LLC
Recall StatusTerminated
Product CodeEFQ
Date Initiated2022-09-28
Date Terminated2026-07-24
Product Quantity1,000 units
Reason for Recall
Specific lots have not been irradiated to eliminate pyronema and therefore the sterility of the affected products may be compromised.
Firm Action
Consignees were notified on about 09/28/2022, via letter. Consignees were instructed to immediately discontinue use and quarantine the products on hand, complete and return the enclosed Acknowledgement Form via fax to 1-855-419-8507 or email to [email protected]. A customer service representative will contact you with a Return Goods Authorization (RGA) number and provide instructions for the return of
products.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Domestic distribution to IL, MS, PA, TN.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.