Terminated

Medacta Shoulder SYSTEM HUMERAL ANATOMICAL METAPHYSIS, REF 04.01.0023 through 04.01.0055, all sizes - Product Usage: The Medacta Anatomic Shoulder Prosthesis is indicated for treatment of humeral fractures and for primary or revision total shoulder replacement in patients with an intact or reparable rotator cuff shoulder joint, severe arthropathy or a previously failed joint replacement.

FDA Recall Z-0775-2020 · Posted None

Recalling FirmMedacta Usa Inc
Recall StatusTerminated
Product CodeKWS
Date Initiated2019-11-04
Date Terminated2021-05-21
Product Quantity207 units

Reason for Recall

Certain lot numbers of the Medacta Humeral Anatomical Cementless Metaphysis have a dimensional condition that could limit insertion of the screw used to lock the metaphysis to the humeral diaphysis.

Firm Action

On 11/4/19 all distributors with recalled product were immediately notified of the recall by phone. On 11/5/19, all distributors were further notified of the recall via email and requested to return their product to the Memphis warehouse.

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution in the states of KY, NJ, IL, SC, CA, TX, ID, AR, OH.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →