Open, Classified

Programmable Diagnostic Computer

FDA Recall Z-0775-2023 · Posted 2023-01-01

Recall StatusOpen, Classified
Product CodeDQK
Date Initiated2022-11-16
Product Quantity2213 distributed worldwide; 638 US

Reason for Recall

The firm will be performing a software update to address a software error which affects the listed products. This correction addresses four potential software issues: 1) "PASSWORD STORE CORRUPTED" error message during system boot; 2) Subsystem crash during examination; 3) Dialog Monitor Computer (DMC) application crash while loading a study; and 4) Software crash due to system internal timeout. Issue 1 may lead to a delay or interruption of procedure. Issues 2, 3, and 4 may result in delay in starting or continuing the examination, and may also prevent the operator from starting or continuing a study;

Firm Action

On November 16, 2022, the firm notified customers with an Urgent Medical Device Correction letter delivered electronically via EchoSign. Customers were informed of the 4 software issues. Siemens will update the software in affected systems to resolve the issues. The firm's service organization will contact customers to arrange a date to perform the corrective action. You may contact the firm's service organization for an earlier appointment at 1--800-888-7436.

Root Cause

Software design

Distribution Pattern

Domestic distribution nationwide. Worldwide foreign distribution to Albania Algeria Angola Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belgium Bolivia Bosnia and Herzegovina Botswana Bulgaria Cambodia Canada Chile China Colombia Costa Rica Croatia Cyprus Czech Republic Denmark Djibouti Dominican Republic Ecuador Egypt El Salvador Fiji Finland France Germany Ghana Greece Guatemala Hong Kong Hungary India Indonesia Iran Iraq Ireland Italy Ivory Coast (Cote d' Ivoire) Jordan Kazakhstan Kenya Kuwait Latvia Lebanon Libya Macedonia Madagascar Malaysia Mauritius Mexico Monaco Mongolia Montenegro Morocco Myanmar Namibia Netherlands New Zealand Nigeria Norway Oman Pakistan Panama Philippines Poland Portugal Qatar Romania Rwanda Saudi Arabia Senegal Serbia Singapore Slovakia Slovenia South Africa South Korea Spain Sudan Sweden Switzerland Syria Taiwan Tanzania Thailand Trinidad and Tobago Tunisia Turkey Turkmenistan Uganda Ukraine United Arab Emirates United Kingdom Vietnam Yemen

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑