Varian brand Clinac 600C (CD, EX) and Portal Vison with Dynamic MLC or RPM system capability

FDA Recall Z-0777-05 · Posted 2005-04-27

Recall StatusTerminated
Product CodeIYE
Date Initiated2005-02-04
Date Terminated2006-08-17
Product Quantity170 units

Reason for Recall

Treatment beam may not completely pause while device changes field shape; Potential for mistreatment could occur.

Firm Action

On 2/4/05, the firm issued letters to all its direct consignees, informing them of the affected product and providing instructions on the device correction.

Root Cause

Software design

Distribution Pattern

The firm has indicated that product was released for distribution to 143 direct consignees in US and to 25 foreign countries via international distribution centers. Foreign distribution include: Australia, Bahrain, Belgium, Canada, Czech Republic, Egypt, Finland, France, Germany, Iceland, India, Israel, Italy, Japan, Kazakhstan, Latvia, Norway, Poland, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, and United Kingdom.