BECKMAN COULTER IRISpec CA/CB/CC control, REF 900-7211, For in vitro diagnostic use only.

FDA Recall Z-0777-2023 · Posted 2023-01-03

Recalling FirmBeckman Coulter, Inc.
Recall StatusOpen, Classified
Product CodeJJW
Date Initiated2022-11-18
Product Quantity3657 units

Reason for Recall

Beckman Coulter has become aware of an increase in customer complaints generated for intermittent IRISpec CB glucose control false positive results when used in conjunction with iChemVelocity Urine Chemistry Strips PN 800-7212

Firm Action

The firm issued a MEDICAL DEVICE RECALL notice to its consignees on 11/18/2022 by mail and email. The notice explained the problem and requested the following actions be taken: "If IRISpec CB controls pass (QC results are within QC acceptance criteria): Continue to process patient samples per IFU (P/N 301-7146BE) If IRISpec CB controls fail (QC results are outside QC acceptance criteria): Refer to Addendum A below for troubleshooting No retrospective review of QC or patient sample results is necessary." "Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter."

Root Cause

Under Investigation by firm

Distribution Pattern

US and Canada, Brazil, Korea, Mexico, Puerto Rico, Taiwan