Medline custom medical procedure kits labeled as: 1) CATH LAB PACK-LF, REF DYNJ0382559M; 2) CRANI, REF DYNJ56001B; 3) GENDER AFFIRMING SURGERY, REF DYNJ66252; 4) GENDER AFFIRMING SURGERY, REF DYNJ66252A; 5) MAJOR NEURO PACK-LF, REF DYNJ0578916X.

FDA Recall Z-0777-2025 · Posted 2024-12-30

Recall StatusOpen, Classified
Product CodeOES
Date Initiated2024-11-21
Product Quantity1236 units

Reason for Recall

Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.

Firm Action

Medline issued a MEDICAL DEVICE VOLUNTARY RECALL notice on 11/21/2024 to its consignees via US Mail and email. The notice explained the issue, potential risk to health, and requested that the affected product be quarantined. Once the response is received, Medline will provide labels to place on affected inventory, with instructions for the consignee to remove the affected component prior to using the kit.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of UAE, and Canada.