American Diagnostica Inc. ACTICLOT dPT Reference 824 in vitro diagnostic
FDA Recall Z-0779-2007 · Posted 2007-05-03
Advertisement
Recalling FirmAmerican Diagnostica, Inc.
Recall StatusTerminated
Product CodeGIR
Date Initiated2007-02-28
Date Terminated2016-07-06
Product Quantity249 kits
Reason for Recall
Stability of product not assured. An increase of clotting time for both normal and abnormal plasmas
Firm Action
American Diagnostica notified accounts by letter on 2/28/07 and 3/02/07 and international distributors on 2/28/07. Users are requested to discontinue use and destroy. Kits will be replaced.
Root Cause
Other
Distribution Pattern
Worldwide, including USA, Canada, Germany, UK, Italy, Israel, Greece, and the Netherlands.