Terminated

Biomet bone mulch screw, 10.5mm x 20 mm, TI-6AL-4V alloy, sterile, Biomet Sports Medicine, Warsaw, IN; REF 907320.

FDA Recall Z-0779-2010 · Posted 2010-02-02

Recalling FirmBiomet, Inc.
Recall StatusTerminated
Product CodeHWC
Date Initiated2009-12-22
Date Terminated2010-10-22
Product Quantity169

Reason for Recall

The screw may not have the hex dimension inside of the head of the screw, which will prevent the device from being implantable. This may cause a delay in the procedure while another screw is obtained.

Firm Action

Hospitals, distributors, and other customers were notified by letter dated 12/22/09 and instructed to locate and return these lots of product. Questions related to this notice should be directed to the firm at 800-348-9500 or 574-372-3983, Monday through Friday, 8AM to 5PM.

Root Cause

Employee error

Distribution Pattern

Arkansas, California, Delaware, Florida, Kansas, Kentucky, Michigan, Missouri, New York, Pennsylvania, South Dakota, Texas, Washington, Wisconsin, Chile, Germany, Italy, Japan and Norway.