Cobalt XT DR MRI SureScan Implantable Cardioverter Defibrillator, product number DDPA2D1
FDA Recall Z-0779-2024 · Posted 2024-01-22
Advertisement
Recalling FirmMedtronic Inc.
Recall StatusOpen, Classified
Product CodeLWS
Date Initiated2023-12-07
Product Quantity1 device
Reason for Recall
During internal review of Polaris battery assembly data from the HiPot operation, a test systems engineer noticed that some batteries had gone through the High Potential test twice. This was able to occur because the test screen did not prevent retesting after a failure. Some batteries were therefore retested and passed.
Firm Action
Medtronic hand delivered an Urgent Medical Device Recall/Retrieval notice to its sole consignee on 12/07/2023. The notice explained the issue and indicated that the device would be collected and a replacement provided.
Root Cause
Nonconforming Material/Component
Distribution Pattern
US Nationwide distribution in the state of Florida.