XiO RTP System The XiO RTP System is used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on-screen and in hard-copy, either two or three-dimensional radiation dose distributions within a patient for a given treatment plan set-up.

FDA Recall Z-0780-2014 · Posted 2014-01-17

Recalling FirmElekta, Inc.
Recall StatusTerminated
Product CodeMUJ
Date Initiated2013-11-25
Date Terminated2014-05-13
Product Quantity13

Reason for Recall

When the number of fractions is changed for a proton spot beam, the MU are recomputed and rounded if necessary, but the dose is not recalculated as it should be even after saving and re-opening the plan.

Firm Action

An Important Safety Notice, LINXIO0017 Proton Sport Beams, Changes in Fractionation not Updating Dose, is in the process of being distributed to affected customers. The release date for this notice is November 25, 2013. The notice included a Confirmation of Receipt which was to be returned to Elekta. For further questions please call (404) 993-5886.

Root Cause

Software Design Change