PerkinElmer 20ul Clear RoboRack Tip, Pipette (Qty 960) Item Number: 6000677. Used with the: JANUS AUTOMATED WORKSTATION Product Codes: AJS4001, AJM4001, AJI4001, AJL4001, AJS8001, AJM8001, AJI8001, AJL8001, AJM4G01, AJI4G01, AJL4G01, AJM8G01, AJI8G01, AJL8G01, AJM4M01, AJI4M01, AJL4M01, AJM8M01, AJI8M01, AJL8M01 NGS EXPRESS Product Codes: AJS4NGS

FDA Recall Z-0780-2017 · Posted None

Recalling FirmPerkinElmer, Inc
Recall StatusTerminated
Product CodeJQW
Date Initiated2016-10-13
Date Terminated2018-09-13
Product Quantity33312

Reason for Recall

Defective tips have the potential to pipette/dispense volumes 50-100% lower than is specified for the tips.

Firm Action

PerkinElmer initiated the recall on Sept 1. 2016 by letter dated . Customers were told of the defect, potential for inaccurate measurements, and requested to inspect for the lots stop using and destroy. Replacement product to be provided. Accounts asked to compelte Response form. Questions call your PerkinElmer representative.

Root Cause

Process control

Distribution Pattern

IA, MD, NY, WA Foreign: France, Germany, The Netherlands, Hungary, Singapore