Terminated

ARCHITECT CA 19-9 Reagent Kit, list 6C04-20 (1 x 100 tests) and 6C04-25 (4 x 100 tests), Abbott laboratories, Abbott Park, IL 60064 U.S.A.

FDA Recall Z-0781-04 · Posted 2004-07-20

Recall StatusTerminated
Product CodeNIG
Date Initiated2004-03-23
Date Terminated2005-04-26
Product Quantity1,987 kits

Reason for Recall

The reagent lots may produce patient results that are lower than expected when compared to other lots of in-date reagents.

Firm Action

Abbott international affiliates were notified of the recall by e-mail on 3/23/04, and requested to notify their customers.

Root Cause

Other

Distribution Pattern

Canada, Australia, Germany and Japan