Terminated

Thyroid Pack CKHA-LF, Model DYNJ50352B - Product Usage: This is a type of convenience kit used in various surgeries including cardiac, abdominal, and neck surgeries.

FDA Recall Z-0781-2021 · Posted None

Recall StatusTerminated
Product CodeOGR
Date Initiated2020-11-13
Date Terminated2021-10-26
Product Quantity40

Reason for Recall

The cylindrical sponge component is not x-ray detectable.

Firm Action

Customers were notified by email on 11/13/2020. Customers are asked to return the response form via fax or email. The recalling firm will send subsequent letters to consignees that have not provided a response to the recall notification within the allotted time. No product is to be returned to Medline Industries. Consignees are instructed to apply labels to their affected kits and return to inventory. The label indicates that there is a recalled component within the pack, not to use it, and replace from a different sterile supply. No other components within the kit are affected.

Root Cause

Process control

Distribution Pattern

Foreign distribution only - Canada.

Official Source

View FDA Recall →