Terminated

TomoTherapy Treatment System - Product Usage: used as an integrated system for the planning and precise delivery of radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissue.

FDA Recall Z-0784-2021 · Posted None

Recalling FirmAccuray Incorporated
Recall StatusTerminated
Product CodeIYE
Date Initiated2020-12-10
Date Terminated2022-09-26
Product Quantity7 units

Reason for Recall

"MLC tickle error" may result in the delivered dose to effectively rotate from the planned dose.

Firm Action

Urgent Medical Device Recall "MLC tickle error" issue during TomoHelical deliver was sent to customers beginning 12/10/2020. Accuray is planning software releases that will address this issue. All TomoTherapy systems at affected software versions will be upgraded to these versions once released.

Root Cause

Software Design Change

Distribution Pattern

Worlwide distribution - US Nationwide distribution including in the states of CA, FL, IL, ND, NJ, OK, UT and the countries of Australia, China, Italy, Switzerland, United Arab Emirates.