Medline custom medical procedure kits labeled as: 1) ENDOSCOPIC SKULL BASE, REF DYNJ58648B; 2) ENDOSCOPIC SKULL BASE, REF DYNJ58648C; 3) ENT ENDO SINUS WILMINGTON PACK, REF DYNJ66447; 4) ENDOSCOPIC SKULL BASE, REF DYNJ58648D; 5) ENDOSCOPIC SKULL BASE, REF DYNJ58648J.
FDA Recall Z-0784-2025 · Posted 2024-12-30
Recalling FirmMEDLINE INDUSTRIES, LP - Northfield
Recall StatusOpen, Classified
Product CodeNWU
Date Initiated2024-11-21
Product Quantity1072 units
Reason for Recall
Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.
Firm Action
Medline issued a MEDICAL DEVICE VOLUNTARY RECALL notice on 11/21/2024 to its consignees via US Mail and email. The notice explained the issue, potential risk to health, and requested that the affected product be quarantined. Once the response is received, Medline will provide labels to place on affected inventory, with instructions for the consignee to remove the affected component prior to using the kit.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of UAE, and Canada.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.