Sunrise Hoyer Advance Patient lifts
FDA Recall Z-0785-05 · Posted 2005-05-03
Advertisement
Recalling FirmSunrise Medical, Inc.
Recall StatusTerminated
Product CodeFSA
Date Initiated2005-02-03
Date Terminated2005-08-08
Product Quantity27 units
Reason for Recall
The leg actuation mechanism on certain Hoyer Advance Patient Lifts may not operate properly, which could cause the lift to become unstable.
Firm Action
Telephone calls were placed to customers on 02/03/05. Urgent Medical Device Correction letters, dated 02/03/05, were also sent to customers.
Root Cause
Other
Distribution Pattern
Domestic distribution: WI, FL, GA, TX, OK. Foreign distribution: Canada. No military or VA facilities
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.