ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems. Distributed by Philips Medical Systems, Cleveland, OH

FDA Recall Z-0785-2014 · Posted 2014-01-17

Recall StatusTerminated
Product CodeIZJ
Date Initiated2013-12-04
Date Terminated2015-11-06
Product Quantity255 units

Reason for Recall

The recalling firm determined that visual inspections of the fork joints behind the detector that are conducted during six month preventive maintenance (PM) may not detect small cracks or deficiencies in the fork weldment.

Firm Action

On October 30, 2013 & December 4, 2013, the firm sent Urgent Medical Device Recall letters to their customers.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution: US (nationwide) including states of: AL, AR, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TN, TX, VA, VT, WA, WI & WV; and Internationally to: Austria, Belgium, Canada, Denmark, Germany, Great Britain, Italy, Netherlands, Norway, Pakistan, Russia, Singapore & South Korea.