Wright Medical Technology, Inc. ORTHOLOC Non-Locking Screw 2mm x 8mm, REF 5201120008, non-sterile, a component of the ORTHOLOC 2.0/2.4 Plate System.

FDA Recall Z-0786-2022 · Posted None

Recall StatusTerminated
Product CodeHRS
Date Initiated2022-02-15
Date Terminated2022-10-05
Product Quantity1 package

Reason for Recall

An incorrect screw is packaged in the product, which contains a locking screw instead of a non-locking screw.

Firm Action

The firm notified the foreign competent authority via email on 2/15/2021 which included a copy of the recall letter. Once the foreign competent authority approves the recall, the recall letter will be sent to the foreign consignee. The recall letter is dated 2/15/2022 and describes the product, the issue, potential risks, and the actions needed. The actions included (1) informing individuals within their organization who need to be aware of this device recall; (2) immediately check all stock areas and/or operating room storage to determine if there is any affected product; (3) quarantine and discontinue use of the recalled device; (4) complete the business reply form and a Stryker representative will contact the customer to arrange for return and replacement of the device; (5) inform Stryker if the device has been distributed to other organizations. If so, provide contact details so that Stryker can inform the recipient appropriately. If the consignee is a distributor, they are responsible for notifying the affected customers; and (6) inform Stryker of any adverse events concerning the use of the device.

Root Cause

Process control

Distribution Pattern

International distribution to the country of France.