Terminated

Multi-Lumen Central Venous Catheterization Kit with Blue Flex-Tip(R), ARROWg+ard Blue PLUS (R) Catheter, Sharps Safety Features and Maximal Barrier Precautions. (contains Q-Syte)

FDA Recall Z-0788-2010 · Posted 2010-03-05

Recall StatusTerminated
Product CodeFOZ
Date Initiated2009-11-24
Date Terminated2012-01-18
Product Quantity180 kits

Reason for Recall

Q-Syte Component contained in kits was recalled by manufacturer due to the potential for embolism if used with a central venous catheter.

Firm Action

The recalling firm issued an Urgent Medical Device Recall letter dated 11/24/09 to their customers informing them of the problem the need to return any kits in inventory and notify their customers of the recall. For additional information customers can contact the Customer Service Department at 800-523-8446.

Root Cause

Nonconforming Material/Component

Distribution Pattern

The products were shipped to distributors in GA, MD, and VA.