LYFO Disk

FDA Recall Z-0788-2022 · Posted None

Recalling FirmMicrobiologics Inc
Recall StatusCompleted
Product CodeJTR
Date Initiated2015-07-17
Product Quantity3 units

Reason for Recall

LYFO Disk Catalog #0983L is to contain Candida tropicalis. However, lot number 983-40-3 under Catalog #0983L, labeled as Candida tropicalis, is actually Yersinia entercolitica. This labeling error may cause a user's quality control to fail, resulting in testing needing to be repeated and delayed test results.

Firm Action

Microbiologics notified customers on 07/17/2015 via letter titled "Medical Device Recall". Instructions included to inspect inventory for affected products. If customers have received the product identified within the recall (LYFO DISK 0983L, Lot# 983-40-3), it is recommended that the customer destroys the product or return it to the recalling firm. Acknowledgement of the receipt of the recall letter was requested by completing a form and returning to Microbiologics. Questions or concerns can be directed to the Quality Systems Manager at 320.229.7050.

Root Cause

Labeling mix-ups

Distribution Pattern

Distribution is nationwide with three consignees located in Colorado, Georgia, and Ohio.