Stryker Xia Precision System; Jam Shidi 10 Gauge 9 inch; Sterile; Manufactured for: Stryker Spine SAS, Cestas, France; Distributed in the US by Howmedica Osteonics Corp, 325 Corporate Drive, Mahwah, NJ. Intended for use as a manual surgical instrument during percutaneous posterior, non-cervical pedicle fixation.
FDA Recall Z-0789-2010 · Posted 2010-02-16
Advertisement
Recalling FirmStryker Spine
Recall StatusTerminated
Product CodeLXH
Date Initiated2010-01-18
Date Terminated2012-09-06
Product Quantity2727 total - all sizes.
Reason for Recall
Stryker Spine received reports of packaging failures during shipping for the Jam Shidi instruments, which may compromise the sterility of the product.
Firm Action
Stryker Spine issued an "Urgent Product Recall" notification dated January 15, 2010 via Federal Express to all consignees. The letter informed users of the issue and potential hazards of the affected product.
For further information, contact your local Stryker Spine Sales Representative or call 201-760-8206.
Root Cause
Package design/selection
Distribution Pattern
Nationwide distribution.