Terminated

Fast-Cath Hemostasis Introducer, HEMOSTASIS INTRODUCER, 8.5F, 2.83 mm, REF 406143 - Product Usage: The Fast-Cath Hemostasis Introducers are indicated for the introduction of angiographic catheters, closed-end catheters, balloon catheters, and electrodes into a vessel where minimizing blood loss is essential.

FDA Recall Z-0789-2020 · Posted None

Recalling FirmAbbott Medical
Recall StatusTerminated
Product CodeDYB
Date Initiated2019-12-09
Date Terminated2020-12-11
Product Quantity380 devices

Reason for Recall

Three lots of Fast-Cath Hemostasis Introducer devices contained 8F sized dilators and sheaths instead of the expected 8.5F devices.

Firm Action

Beginning on December 9, 2019, US consignees will be notified in person by an Abbott field representative. OUS consignees will be notified by mail. The firm is seeking return of the devices.

Root Cause

Packaging

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of MA, VA, NH, FL, CT, UT, OH, WI, NY, IA, MI, MS, PA, and countries of Canada, Czech Republic

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.