Guidant EasyTrak 8 French Guiding Catheter, Inner Diameter .087'', 2.21 mm, Overall Length 54 cm, CS Wide Tip Shape, CS-W, REF 6705
FDA Recall Z-0790-04 · Posted 2004-07-20
Advertisement
Recalling FirmGuidant Corp-Cpi Division
Recall StatusTerminated
Product CodeDQR
Date Initiated2004-03-11
Date Terminated2005-10-31
Product Quantity213 catheters
Reason for Recall
The terminal hub component in a some of the catheters has an unintended abrupt transition between the entry port and the catheter lumen which can result in difficulty introducing or removing delivery system components during lead placement.
Firm Action
Starting on March 11, 2003, the firm''s representatives visited consignees, to whom the recalled catheters had been consigned, and retrieved the catheters.
Root Cause
Other
Distribution Pattern
Nationwide in the United States.