ImplantDirect InterActive Precision IO Scan Adapter, Part #6530-09PT, Platform: 3.0mm, Non-Sterile, Rx. The firm name on the label is Implant Direct Sybron Manufacturing LLC, Thousand Oaks, CA.

FDA Recall Z-0791-2020 · Posted None

Recall StatusCompleted
Product CodeNDP
Date Initiated2018-11-28
Product Quantity232 devices

Reason for Recall

The peek part was loose on the titanium base of the device which could result in it falling into a patient's mouth during a procedure resulting in aspiration or ingestion.

Firm Action

The firm first issued recall letters on 11/28/2018 via FedEx for Part number 6534-09PT and then on 12/28/2018, issued recall letters for Part number 6530-09PT via FedEx. On 2/14/2019, the firm issued another letter to their consignees that added a risk statement regarding the recall of both part numbers (6534-09PT and 6530-09PT).

Root Cause

Process control

Distribution Pattern

Distribution was made to AL, AZ, CA, FL, GA, KS, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, and VA. There was no government/military distribution. Foreign distribution was made to Canada, Armenia, Australia, Denmark, France, Germany, Great Britain, Iceland, Italy, Lithuania, Netherlands, Poland, Spain, Sweden, and Switzerland.