Medline custom medical procedure kits labeled as: 1) ENSEMBLE RACHIS-LF, REF DYNJ47568B; 2) UROLOGY ROBOTIC, REF DYNJ55555D; 3) KIT,SPINE LH, REF DYNJ906448A.

FDA Recall Z-0792-2025 · Posted 2024-12-30

Recall StatusOpen, Classified
Product CodeOHD
Date Initiated2024-11-21
Product Quantity2664 units

Reason for Recall

Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.

Firm Action

Medline issued a MEDICAL DEVICE VOLUNTARY RECALL notice on 11/21/2024 to its consignees via US Mail and email. The notice explained the issue, potential risk to health, and requested that the affected product be quarantined. Once the response is received, Medline will provide labels to place on affected inventory, with instructions for the consignee to remove the affected component prior to using the kit.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of UAE, and Canada.