RadSuite, a picture archive and communications system. Model Numbers Versions 5.30.0, 5.30.1, 5.30.2, 5.30.2 SB0344, 5.30.3, 5.30.4, 5.30.4 HF0418E, 5.30.5, 5.30.5 HF0404E, 5.30.5 HF0413E, 5.30.5 HF390E, 5.30.5 SB0368, 5.30.6, 5.30.6 HF0405E, 5.30.6 HF0406E, 5.30.7, 5.30.7 HF0429E, 5.30.7 HF0441E, 5.30.7 HF0446E, 5.30.7 HF0470E, 5.30.7 HF0473E, 5.30.8, 5.30.8 HF0477E, 5.30.8 HF0490E, 5.30.8 HF0495 SB0508E, 5.30.8 HF0495E, 5.30.8 HF0495E SB504E, 5.30.8 HF0550, 5.30.8 SB0485E, 5.30.8 SB0486E, 5.30.X SB0381E, 5.35.1, 5.35.1 HF0487E, 5.35.1 HF0489E, 5.35.2, 5.35.3, 5.35.4, 5.35.4 HF0513, 5.35.4 HF0518, 5.35.4 HF0528, 5.35.4 HF0531, 5.35.4 HF0535, 5.35.4 HF0546, 5.35.4 HF0555, 5.35.4 HF0555.1, 5.35.4 HF0555.2, 5.35.4 HF0555.3, 5.35.4 HF0560, 5.35.4 Powerscribe Validation, 5.35.5, 5.35.5 From CCS, 5.35.5.1, 8.30.0, 8.30.1, 8.30.3, 8.30.3.1, 8.30.4, 8.30.5, 8.30.6, 8.30.6.1, 8.30.6.2, 8.30.6.3, 8.30.7, 8.30.7.1, 8.30.7.2, 8.30.7.3, 8.30.7.3-Hopkins, 8.30.7.4, 8.30.7.5, 8.30.7.5.b, 8.30.7.6, 8.30.7.7, 8.30.7.8.

FDA Recall Z-0794-2016 · Posted None

Recall StatusTerminated
Product CodeLLZ
Date Initiated2015-12-17
Date Terminated2016-05-10
Product Quantity164

Reason for Recall

Potential incorrect Standardized Uptake Values (SUV) measurements in RadSuite.

Firm Action

Merge Healthcare sent an " Urgent - Medical Device Recall letter dated January 21, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer . If you may have further distributed this product, please identify your customers and notify them at once of this product recall. Your notification to your customers may be enhanced by including a copy of this recall notification letter. Merge Healthcare is committed to improve efficiencies and enhancing the quality of healthcare worldwide. If you have any additional questions, call Merge Customer Service at (877) 741-5369 or [email protected].

Root Cause

Software design

Distribution Pattern

Nationwide Distribution to the states of : MN, WI, IL, PA, AL, MO, MT, AZ, CA, OH, FL, SC, MI, TN, CO, NV, IN, NY, MD, AR, KS, NY, IN, IA, TX, OK, NC, WA and DC.