Terminated

Ascension Modular Radial Head (MRH) Radial Head Implant, Catalog Number, MRH-350-04-WW

FDA Recall Z-0794-2019 · Posted None

Recall StatusTerminated
Product CodeHSD
Date Initiated2018-11-16
Date Terminated2019-11-22

Reason for Recall

The product was packaged with the Instructions for Use for a different product (Ascension First Choice DRUJ System Partial Ulnar Head Implant).

Firm Action

On 11/16/18, the recalling firm sent out an Urgent Voluntary Medical Device Correction letter via email to affected customers to notify them of the product issue. A second, updated letter was sent on 12/13/18. The firm requests that all recipients please examine their inventory to determine if they have any MRH Radial Head Implants, catalog number, MRH-350-04-WW, from lot number, 170691. Customers: - If you do have any MRH Radial Head Implants, catalog number, MRH-350-04-WW, from lot number, 170691 please discard the First Choice DRUJ System Partial Ulnar Head Implant instructions packaged with the product and use the attached correct IFU provided to you along with this communication letter. Distributors: - If you do have any unused MRH Radial Head Implants, catalog number, MRH-350-04-WW, from lot number, 170691 please return them to Integra. - We also kindly ask you to provide us the final customer list who may have the affected products. Integra will provide them the correct IFU. Complete the attached Acknowledgement Form and return the completed form by email or fax indicating receipt and review of this notification. Keep a copy of the form for your records. Should you have any questions regarding these instructions, please contact Customer Service Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: [email protected] Outside USA: 1- 609-936-5400; email: [email protected]

Root Cause

Vendor change control

Distribution Pattern

Distributed to accounts in MO. Foreign distribution to Korea and Thailand.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.