Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

AEROSET System External Waste Pump

FDA Recall Z-0796-03 · Posted 2003-05-09

Recall StatusTerminated
Product CodeJJE
Date Initiated2003-01-27
Date Terminated2004-04-07
Product Quantity461 units

Reason for Recall

External waste pump has metal edge around perimeter sharp enough to cause scratch or cut.

Firm Action

Firm sent 01/27/2003 letter to all consingees informing them of the issue. US Cutomer Reply form will be distributed with the customer letter.

Root Cause

Other

Distribution Pattern

Nationwide and to the following foreign countries: Mexico, Chilie, Columbia, Dominican Republic, Hong Kong, Guatemala, Argentina, Canada, Australia, Venezuela, Brazil, Germany, New Zealand, Malaysia, Japan, Taiwan, Korea, Singapore, Thailand

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑