Terminated

Altaire, Magnetic Resonance Imaging Device.

FDA Recall Z-0797-05 · Posted 2005-05-03

Recall StatusTerminated
Product CodeLNH
Date Initiated2005-04-11
Date Terminated2012-04-12
Product Quantity212 devices

Reason for Recall

The electrical cables that supply gradient power to the magnet can become loose over time and create a potential fire hazard.

Firm Action

The firm sent a recall/field correction letter, dated 4/11/05, to direct accounts.

Root Cause

Other

Distribution Pattern

Direct accounts located in AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI.