FAST-FIX AB Curved Needle Delivery System, Used for meniscal repair. Product Number: 7209399

FDA Recall Z-0798-06 · Posted 2006-04-26

Recall StatusTerminated
Product CodeKGS
Date Initiated2006-03-24
Date Terminated2007-12-12
Product Quantity1800 units

Reason for Recall

The implant consists of a T1 and T2 anchor which are joined together by a suture. The T2 portion of the implant may not advance into the deployment position preventing deployment of the T2 anchor.

Firm Action

Smith & Nephew notified accounts by letter dated 3/24/06 issued by Federal Express. Users are requested to return product.

Root Cause

Other

Distribution Pattern

Nationwide Foreign: Austria, Germany, France, Italy, Australia