FAST-FIX AB Curved Needle Delivery System, Used for meniscal repair. Product Number: 7209399
FDA Recall Z-0798-06 · Posted 2006-04-26
Advertisement
Recalling FirmSmith & Nephew, Inc. Endoscopy Division
Recall StatusTerminated
Product CodeKGS
Date Initiated2006-03-24
Date Terminated2007-12-12
Product Quantity1800 units
Reason for Recall
The implant consists of a T1 and T2 anchor which are joined together by a suture. The T2 portion of the implant may not advance into the deployment position preventing deployment of the T2 anchor.
Firm Action
Smith & Nephew notified accounts by letter dated 3/24/06 issued by Federal Express. Users are requested to return product.
Root Cause
Other
Distribution Pattern
Nationwide Foreign: Austria, Germany, France, Italy, Australia
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.