Porous titanium flared rim acetabular shell with three clustered holes in even sizes from 48 to 66mm, a hip prosthesis component. The lot number on the product labeling is the unique identifier of this component.

FDA Recall Z-0799-06 · Posted 2006-04-27

Recalling FirmEncore Medical, Lp
Recall StatusTerminated
Product CodeLPH
Date Initiated2006-01-12
Date Terminated2006-07-11
Product QuantityTotal of 30 units

Reason for Recall

Voids in the porous coating of a component used in total hip replacement surgery.

Firm Action

Firm notified all sales agents by letter on January 12, 2006 with instructions to return the product. The second letters went out on January 30, 2006 to the sales associates with instructions to contact the implanting physicians.

Root Cause

Other

Distribution Pattern

Nationwide. Iowa, Florida, California, Arizona, South Dakota, Utah, Missouri, Texas, and Massachusetts, Maryland, Oklahoma, Nevada