Sheridan/CF Cuffed Tracheal Tube, Sterile, Rx only, Teleflex Medical, Research Triangle Park, NC 27708. Designed for oral and nasal intubation and are indicated for airway management.
FDA Recall Z-0799-2015 · Posted 2014-12-12
Advertisement
Recalling FirmTeleflex Medical
Recall StatusTerminated
Product CodeBTR
Date Initiated2014-11-12
Date Terminated2017-10-13
Product Quantity14,000 units
Reason for Recall
Product Labeling; Units from the affected lot indicates that it contains a size 8mm tracheal tube but may contain a size 7.5mm tracheal tube.
Firm Action
Consignees were notified by an Urgent Medical Device Recall Notification letter on 11/12/2014. The letter identified the affected product, as well as, the reason for the recall. Customers are instructed to immediately discontinue use of and quarantine any affected product. They are also instructed to complete and fax or e-mail the enclosed Recall Acknowledgement Form regardless of whether they have affected product in stock or not. Upon receipt of the form, a customer service representative will contact customers to provide an RGA number, as well as, instructions for the return of the product. A local sales representative or customer service should be contacted with questions at 1-866-246-6990.
Root Cause
Process control
Distribution Pattern
Worldwide Distribution -- U.S., Belgium, Canada, and Japan.