Gridlock Ankle Screw Driver Bit REF 320-35-003 - Product Usage: A screwdriver facilitates implant insertion and removal.

FDA Recall Z-0799-2020 · Posted None

Recall StatusTerminated
Product CodeHXX
Date Initiated2019-03-14
Date Terminated2022-03-16
Product Quantity20 units

Reason for Recall

A screwdriver used to facilitate implant insertion and removal was found to be out of specification for the minor diameter of the drive feature causing the driver to be unable to retain screws as intended.

Firm Action

The recall strategy included sending written recall communication and response forms as of 03/14/2019 via FedEx (Express Standard Overnight) to all consignees that showed possession / had been sold a device (as the device is considered reusable) as of 03/14/2019. The communication instructed the consignees to: " Check inventory stock for identified devices " Locate and remove and/or discontinue use of identified devices " Return all related product(s) to Recalling Firm in accordance with the included return response / acknowledgement form.

Root Cause

Process control

Distribution Pattern

US Nationwide distribution in the states of CO, CT, IA, TX and WI. OUS: None

Official Source

View FDA Recall →