Recall StatusTerminated
Date Initiated2019-11-27
Date Terminated2024-05-28
Product Quantity27168
Reason for Recall
If the venous cannula is used in an unintended manner (longer than 6 hours; particularly during extracorporeal membrane oxygenation [ECMO]) there is a risk that separation of the cannula from its connector can occur; which can result in significant blood loss.
Firm Action
On 11/27/19, Urgent Medical Device Safety Notices were mailed to customers (Attention: Risk Management Departments). Customers were informed the following:
1. Review the field safety notice to understand the potential hazard.
2. Meet and review with the appropriate clinical staff at your hospital.
3. Complete and return the acknowledgment form attached to this letter via fax (949-250-2525).
4. Distribute this notice within your organization or to any organization where the devices under the safety alert have been transferred.
5. Make the information available to personnel that may be using these devices as part of continuing communication and training.
6. Instructions for Use are being modified to add a contraindication for long term use (greater-than 6 hours, particularly during extracorporeal membrane oxygenation (ECMO) procedures.
7. Product return is not required.
If you have additional questions, please contact customer service at 800-424-3278, Option 1, from 5:00 AM - 4:30 PM PST.
Distribution Pattern
U.S.: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV.
O.U.S. (Foreign): Brazil, Colombia, Albania, Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Czech Republic, Denmark, France, Germany, Greece, Ireland, Israel, Italy, Japan, Jordan, Malaysia, Namibia, Poland, Romania, Portugal, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom.