Sentinel Implantable Carioverter-DefibrilIator (ICD) model 2012 is a multi-programmable, two-zone, tiered-therapy device that incorporates anti-tachycardia pacing (ATP), cardioversion, and defibrillation (DEF) tachyarrhythmia therapies. The device also includes a programmable Hot Can electrode and bradycardia pacing support.
FDA Recall Z-0802-05 · Posted 2005-05-06
Advertisement
Recalling Firmela Medical Llc
Recall StatusTerminated
Product CodeLWS
Date Initiated2005-01-25
Date Terminated2006-07-22
Product Quantity3 in US and 2 OUS
Reason for Recall
A change in the automatic capacitor reform schedule during the later portion of the device's useful life can result in excessive 750V charge times and disable the device's ability to report an End of Life (EOL) indicator due to excessive charge times unless the patient is receiving periodic 750V therapy shocks.
Firm Action
Angeion notified physicians responsible for managing each remaining implanted ICD, also provided them with programming means to change the automatic cap charging schedule, and to monitor and document the reprogramming process.
Root Cause
Other
Distribution Pattern
AZ, CA, GA, IL, IN, KY, MA, MD, MI, MN, NC, NJ, OH, PA and to UK, Italy, Germany and Argentina.