The 2nd Assist Knee Positioner
FDA Recall Z-0803-2018 · Posted None
Advertisement
Recalling FirmZimmer Biomet, Inc.
Recall StatusTerminated
Product CodeJEB
Date Initiated2017-09-07
Date Terminated2018-08-23
Product Quantity38,879 units
Reason for Recall
Zimmer Biomet is conducting a medical device recall for the 2nd Assistant Knee Positioner due to the potential for the sterile packaging barrier to be breached during distribution.
Firm Action
On September 12, 2017 an URGENT MEDICAL DEVICE RECALL notice was issued to distributors and customers titled "The Second Assistant Disposable Knee Positioner". The notice requests customers to review their inventory and to quarantine any remaining product, which will be returned to the firm. Also, to notify any additional accounts which may have received product from you. Questions or concerns can be directed to Customer Service at 574-371-3071
Root Cause
Package design/selection
Distribution Pattern
Nationally