e-fix E20 power drive. The product is a wheelchair component/accessory.
FDA Recall Z-0804-05 · Posted 2005-05-06
Advertisement
Recalling FirmFrank Mobility Systems, Inc.
Recall StatusTerminated
Product CodeITI
Date Initiated2004-08-01
Date Terminated2006-08-23
Product Quantity314 units
Reason for Recall
may short circuit within the interface due to long term exposure of liquid
Firm Action
The recalling firm issued a Corrective Action Notice to their customers on 9/1/04 via certified mail return receipt. The notice described the problem and the correction (interface was designed). The notice also instructed the dealers/distributors to notify their customers and have the units brought back to them for the exchange.
Root Cause
Other
Distribution Pattern
The products were shipped to distributors nationwide. The product was also shipped to Mexico and Panama.