BREG Sterile Polar Pads; 02356 Rev M, PAD M/U XL STER, Mfg.Date: 112012. Model #s: 02510, 09901, 02330, 02350, 02490, 02356, 02496, 02410, 02430. Local anesthetic effect.
FDA Recall Z-0804-2013 · Posted 2013-02-12
Advertisement
Recalling FirmBreg Inc
Recall StatusTerminated
Product CodeILO
Date Initiated2013-01-14
Date Terminated2013-07-01
Product Quantity44,883 units
Reason for Recall
The recall was initiated because Breg has determined that some Sterile Polar Pads products manufactured from January 2010 to October 2012 may have sustained damage to the product packaging which may compromise product sterility assurance.
Firm Action
The firm, BREG, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated January 14, 2013 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately examine your inventory and quarantine the product; complete the Return Response Form by email, Fax or mail; and contact Breg for a Return Authorization for the affected product and request replacement by contacting Breg Customer Care at 800-321-0607. Note: response is required even if you have no affected inventory.
Should you have any questions regarding this communication or need to report an adverse event, please contact Breg Customer Care at 800-321-0607.
Root Cause
Packaging
Distribution Pattern
Worldwide Distribution-USA (nationwide) and the countries of Australia, Singapore, Chile, Latvia.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.