Radiomat M+ NIF 14 x 17, Radiomat M+ NIF 35CM x 43CM, Radiomat M+ NIF 10 x 12, Radiomat M+ NIF 25,4CM x 30,5CM
FDA Recall Z-0804-2015 · Posted 2014-12-16
Advertisement
Recalling FirmAGFA Healthcare Corp.
Recall StatusTerminated
Product CodeIWZ
Date Initiated2014-10-24
Date Terminated2015-03-16
Product Quantity142 units
Reason for Recall
Some material of the coating Type CXCPMV3 has potential artifacts showing a small bright thin line when the film is processed.
Firm Action
Dealers were contacted by Agfa HealthCare via phone and/or email on 10/24/2014 and the week of 11/10/14 to inform them of the quality issue and to quarantine identified product shipped to their facility or end users. On 11/17/2014, an "URGENT SAFETY NOTICE" letter was sent to the US dealers. The letter described the safety alert and the request to quarantine affected product. Acknowledgment via FAX-Back or email that the information was received and understood was requested from the dealers. A customer Urgent Field Safety Notice was also provided to those dealers who had shipped affected product to end users so they could sent it to the end use customers.
Root Cause
Other
Distribution Pattern
US Nationwide Distribution in the states of: AL, AZ, CA, LA, MI, NE, NH, NJ, NC, VA
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.