CoaguChek XS PST Meters- IVD (Patient Self-Testing) measures blood-clotting time for people who are taking anticoagulation medications such as Coumadin or warfarin. Catalog 04837738001

FDA Recall Z-0804-2021 · Posted 2021-01-07

Recall StatusTerminated
Product CodeGJS
Date Initiated2020-12-02
Date Terminated2022-06-30
Product Quantity716,880 units

Reason for Recall

Possible Misinterpretation of Displayed Measuring Units from INR to %Quick or Seconds; providing instructions to users for ensuring the meter is displaying INR results

Firm Action

Roche issued on 12/2/2020, UMDC letter notifications via UPS Ground (signature required upon receipt) to Healthcare Professionals, Distributors and Patients provides instructions to users for ensuring the meter is displaying INR results as follows: For the CoaguChek XS Pro system: UMDC for Healthcare Professionals (TP-01147); UMDC TP-01150 (Distributors), including an enclosed UMDC for Healthcare Professionals (TP-01147). For the CoaguChek XS PST system: UMDC for IDTFs (TP-01148) Including an enclosed UMDC for Patients (TP-01142);" UMDC for Clinics (TP-01149);o Including an enclosed UMDC for Patients (TP-01142). The UMDCs posted to the diagnostics.roche.com website. Distributors of CoaguChek Products are instructed as follows: " Read the enclosed Healthcare Professionals UMDC TP-01147 " Within 14 days from receipt of this UMDC distribute a copy of the enclosed Healthcare Professionals UMDC TP-01147 to all customers to whom you have shipped test strips to use with the Coaguchek XS Professional meter within the last 2 years. " If your organization is contractually authorized by Roche to ship test strips to use with the CoaguChek XS Professional meter to redistributors, then your organization must immediately send a copy of this Distributor UMDC TP-01150 and a copy of the enclosed Healthcare Professional UMDC TP-01147 to those re-distributor customers upon your receipt of this UMDC. Such redistributors should take each of the steps included in this Actions Required section of this Distributor UMDC with respect to their own healthcare professional customers. " Complete all sections of the enclosed fax back form (TP-01153); fax or email it according to the instructions on the form. " File this Urgent Medical Device Correction (UMDC) and the enclosed Healthcare Professional UMDC (TP-01147) for future " Healthcare Professionals (HCPs) considered direct government customers (i.e. VA, other government accounts) " If testing a patients INR at your f

Root Cause

Device Design

Distribution Pattern

US Nationwide