ORCHESTRA/ORCHESTRA PLUS Programmer
FDA Recall Z-0805-2014 · Posted 2014-01-22
Advertisement
Recalling FirmSorin Group Italia S.r.l.
Recall StatusTerminated
Product CodeLNX
Date Initiated2013-11-20
Date Terminated2014-09-08
Product Quantity1,718
Reason for Recall
Sorin has voluntarily issued a notification to physicians related to the overestimation of the residual longevity displayed by its programmer (ORCHESTRA or ORCHESTRA PLUS) during a follow-up exam of patients implanted with a REPLY or ESPRIT pacemaker.
Firm Action
On Nov 20 and Nov 25, 2013, the Dear Doctor
Letter was sent by certified mail to 1,310 physicians who are following 13,003 patients
implanted with a REPLU or ESPRIT pacemakers in the US. Sorin provided the following recommendations to the physicians:
You should consider checking the battery impedance of the last follow-up exam. In case
the battery impedance is greater than or equal to 3.5k, a follow-up visit must be
scheduled within a maximum of 6 months from the last follow-up visit.
When pacemaker operation is checked by the simple application of a magnet, a magnet
rate less than 95 min-1 should trigger a follow-up exam in the pacemaker centre.
As a general rule, a maximum of 6 month follow-up interval when the battery impedance
becomes greater than or equal to 3.5 k. This recommendation should also be followed
subsequent to the installation of the new programmer software version.
Root Cause
Software design
Distribution Pattern
Nationwide.