Endotek Alimaxx-B Uncovered Biliary Stent for Endoscopic Delivery, REF 70129-210, 10 x 80, Sterile EO, Merit Medical Systems, Inc., South Jordan, Utah 84095. Intended for palliation of malignant neoplasms in the biliary tree.

FDA Recall Z-0806-2010 · Posted 2010-02-22

Recall StatusTerminated
Product CodeFGE
Date Initiated2010-01-28
Date Terminated2010-04-27
Product Quantity51 units

Reason for Recall

Degradation of the outer portion of the delivery catheter may prevent proper deployment of the biliary stent.

Firm Action

Merit Medical Systems, Inc initiated customer notification by phone and letter beginning January 28, 2010. Consignees were instructed to immediately contact their affected customers, advise them of the recall activity and to identify and quarantine any unused inventory. A Recall Notification Form is to be completed and signed by the sales rep and by the site representative. Affected product is to returned to the firm. For further information, contact Merit Medical Systems, Inc. at 1-801-208-4228.

Root Cause

Component change control

Distribution Pattern

Nationwide Distribution -- AL, CA, CO, FL, IL, KY, NY, MA, MD, ME, MN, MO, PA, TX, UT, VA, WI, VA, TX and WI.