Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile.

FDA Recall Z-0808-2025 · Posted 2025-01-06

Recalling FirmAerin Medical, Inc.
Recall StatusOpen, Classified
Product CodeGEI
Date Initiated2024-11-12
Product Quantity479 units

Reason for Recall

Some units in a single production lot were programmed incorrectly which will result in the VivAer Stylus being incorrectly recognized as RhinAer Stylus when connected to the Aerin Console.

Firm Action

The recalling firm issued emails on 11/12/2024. It was necessary to notify several consignees by mail with an "URGENT: MEDICAL DEVICE FIELD NOTICE" letter dated 11/20/2024. The email described the product purchased by the consignee and the reason for recall. The consignee is informed the programmed treatment time will be shorter than the recommended 30 second treatment time for VivAer and the use of the device may result in insufficient treatment. Under the heading "Action to be Taken," it says if a defective device is identified or the consignee wishes to return unused devices within this production lot, they are to return them to the recalling firm, and return instructions are provided. Regarding the mailed "URGENT: MEDICAL DEVICE FIELD NOTICE" letters dated 11/20/2024, they informed the consignee that some of the units in the production lot were programmed incorrectly and, in these cases, the programmed treatment time will be 12 seconds of radiofrequency energy delivery with no cooling time, not the 30 second treatment time for VivAer that includes 18 seconds of energy delivery and 12 seconds of cooling time. The letter also contains instructions to share this notification with all device users within their facility and network to ensure they are aware of this issue. This notification should also be shared with any organization where the potentially affected devices have been transferred. If you need any further information or support concerning this notification, please contact us at [email protected] or call 512-221-9956.

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide Distribution: AL, AZ, CA, CO, GA, ID, IL, IN, KY LA, MI, MN, NY, NE, NY, OH, OR PA, SC, TN, TX, UT, VA, WA, WI, and WV.