Medtronic Micro Jewel II Implantable Cardioverter Defibrillators, Model 7223Cx

FDA Recall Z-0811-04 · Posted 2004-07-20

Recall StatusTerminated
Product CodeLWS
Date Initiated2004-04-02
Date Terminated2005-11-29
Product Quantity5839 units (2394 units in the United States)

Reason for Recall

Some of the devices with suspect capacitors (high voltage capacitors associated with prior notifications in 1999 and in 2000) have had unexpected charge circuit time-outs or charge circuit inactive conditions as the battery voltage nears the level for Elective Replacement of the devices.

Firm Action

A letter, dated April 2004, was sent on April 5, 2004 to physicians following patients who had the defibrillators implanted. The letter gave recommendations to physicians for selective replacement of some of the implanted devices and for avoiding the problem. Physicians are being visited by sales representatives to verify notification.

Root Cause

Other

Distribution Pattern

Nationwide in the United States and worldwide.