MX441B, CVP Manometer. The product is packaged in a cardboard box. The units are packed ten (10) per case. The product is labeled on the outer case with a stick-on label that reads in part, ''**medex Dublin, OH 43016 USA**STERILE***MX441B CVP Manometer**Lot#***Latex Free''
FDA Recall Z-0812-06 · Posted 2006-04-26
Recalling FirmMedex Inc
Recall StatusTerminated
Product CodeKRK
Date Initiated2005-11-18
Date Terminated2007-05-07
Product Quantity6910 devices
Reason for Recall
The product is labeled as 'Latex Free' , however,one of the components of the finished product may contain latex.
Firm Action
The firm sent a recall letter dated 11/18/2005, via Certified mail to all of their customers. In the letter, the firm requests that the consignee(s) telephone the firm and obtain a Return Authorization number in order to return the recalled product.
Root Cause
Other
Distribution Pattern
The recalled device was distributed to customers throughout the U.S.A. and in Canada.