STA Compact and STA Compact CT For in vitro Diagnostic Use Only In vitro diagnostic analyzer Diagnostica Stago France
FDA Recall Z-0812-2007 · Posted 2007-05-16
Advertisement
Recalling FirmDiagnostica Stago, Inc.
Recall StatusTerminated
Product CodeGKP
Date Initiated2007-01-18
Date Terminated2007-12-10
Product Quantity693 units
Reason for Recall
Result of complaints of field failures of the silicon bridge rectifier, resulting in smoke billowing out of the unit and burnt circuit boards, wires and connectors.
Firm Action
A recall notification has not been sent. The firm did issue a technical bulletin to their representatives.--Update: based on FDA input the firm issued a Recall Letter to all users dated May 25, 2007
Root Cause
Other
Distribution Pattern
Product was distributed to hospitals and laboratories nationwide.