CoaguChek Pro DM meter; catalog # 1894048 and 1893211.
FDA Recall Z-0813-03 · Posted 2003-05-14
Advertisement
Recalling FirmRoche Diagnostics Corp.
Recall StatusTerminated
Product CodeJPA
Date Initiated2003-03-12
Date Terminated2005-02-03
Product Quantity1322
Reason for Recall
High frequency signals from other medical or communications equipment will result in a variable positive bias in test results.
Firm Action
Urgent Product Removal letters dated 3/12/03 were sent to all customers. Distributors were asked to conduct a recall from their customers. Replacement meters are not currently available from Roche. Roche will temporarily provide a CoaguChek S system to run PT samples and will reimburse for APTT and ACT tests that are sent to an accredited outside laboratory.
Root Cause
Other
Distribution Pattern
United States
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.