Abbott i-STAT ACT Kaolin Cartridges. In vitro diagnostic test which uses fresh whole blood to monitor high-dose heparin anticoagulation frequently associated with cardiovascular surgery.

FDA Recall Z-0814-2010 · Posted 2010-02-23

Recall StatusTerminated
Product CodeJBP
Date Initiated2009-08-18
Date Terminated2010-05-18
Product Quantity1,110,775 Cartridges

Reason for Recall

Labeling discrepancy: The manual instruction state that the line should be flushed with 5 mL of saline and the first 5 mL of blood or six dead space volumes should be discarded. The Cartridge Test Information sheets that it should be 2 mL of blood - which is incorrect.

Firm Action

Abbott Point of Care issued an "Important Product Information" letter beginning August 18, 2009. Consignees were instructed on how to reconcile all affected product. For further information, contact Abbott Point of Care Technical Support at 1-800-366-8020, Option 1.

Root Cause

Labeling False and Misleading

Distribution Pattern

Worldwide distribution.